Rem. |
Parameter |
Description |
Minimum |
Maximum |
Default |
Unit |
| S__ | | | Functions about cycle | | | | |
| | St_ | | Functions about cycle temperature | | | | |
| | | _t0 | final temperature | -55.0 | 145.0 | 2.0 | °C |
1 | | | _td | final temperature differential | 0.0 | 50.0 | 0.0 | K |
| | Sd_ | | Functions about cycle timing | | | | |
2 | | | Sd0 | maximum cycle duration | 0 | 194 4:20:15 | 2:00 | dd hh:mm:ss |
| c__ | | | Functions about door and light | | | | |
| | cP_ | | Door switch and evaporator fan | | | | |
| | | cPH | stop fan when door is open | oFF | _on | _on | / |
| | | cPF | pause cycle timer when door is open | oFF | _on | _on | / |
| | | cPd | delay of fan automatic switch on | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | cI_ | | Functions about light | | | | |
| | | cIH | switch on the light when the door is open and off when closed | oFF | _on | _on | / |
| | | cIo | switch off the light automatically if it has been switched on from outside | oFF | _on | _on | / |
| | | cId | delay of light automatic switch off | 0 | 194 4:20:15 | 30 | dd hh:mm:ss |
| b__ | | | Functions about probe calibration | | | | |
| | b1_ | | Probe nr. 1 | | | | |
| | | b1C | room temperature | -9.0 | 9.0 | 0.0 | K |
| | | b1A | enable probe | oFF | _on | _on | / |
| L__ | | | Functions about alarm and stand-by | | | | |
| | Lt_ | | Temperature alarm | | | | |
3 | | | LtL | low temperature alarm set point | -55.0 | 145.0 | -2.0 | °C |
4 | | | LtH | high temperature alarm set point | -55.0 | 145.0 | 14.0 | °C |
| | | Ltd | alarm delay | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | LO_ | | Door alarm | | | | |
| | | LOH | enable door alarm | oFF | _on | _on | / |
| | | LOd | door alarm delay | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | | LOt | temperature alarm minimum delay after door opening | 0 | 194 4:20:15 | 15:00 | dd hh:mm:ss |
| | Lo_ | | On / stand-by status | | | | |
| | | Loo | actual status: stand-by or on | oFF | _on | oFF | / |
| I__ | | | Functions about input-output and machine state (read only) | | | | |
| | IA_ | | Analog inputs | | | | |
| | | IA1 | room temperature | -55.0 | 145.0 | -55.0 | °C |
| | Id_ | | Digital input | | | | |
| | | Id4 | door closed | oFF | _on | oFF | / |
| | OS_ | | Machine status | | | | |
| | | OSn | fan stopped by door opening or manual control | oFF | _on | oFF | / |
| | | OSd | time left to complete the cycle | 0 | 194 4:20:15 | 0 | dd hh:mm:ss |
| | OA_ | | Analog output | | | | |
| | | LLA | actual alarm - read only (0 means no alarm) | 0 | 255 | 0 | / |
| | Od_ | | Digital output | | | | |
5 | | | Od1 | fan | oFF | _on | oFF | / |
| | | Od7 | light | oFF | _on | oFF | / |
| S__ | | | Functions about cycle | | | | |
| | St_ | | Functions about cycle temperature | | | | |
| | | _t0 | final temperature | -55.0 | 145.0 | 2.0 | °C |
1 | | | _td | final temperature differential | 0.0 | 50.0 | 0.0 | K |
| | Sd_ | | Functions about cycle timing | | | | |
2 | | | Sd0 | maximum cycle duration | 0 | 194 4:20:15 | 2:00 | dd hh:mm:ss |
| c__ | | | Functions about door and light | | | | |
| | cP_ | | Door switch and evaporator fan | | | | |
| | | cPH | stop fan when door is open | oFF | _on | _on | / |
| | | cPF | pause cycle timer when door is open | oFF | _on | _on | / |
| | | cPd | delay of fan automatic switch on | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | cI_ | | Functions about light | | | | |
| | | cIH | switch on the light when the door is open and off when closed | oFF | _on | _on | / |
| | | cIo | switch off the light automatically if it has been switched on from outside | oFF | _on | _on | / |
| | | cId | delay of light automatic switch off | 0 | 194 4:20:15 | 30 | dd hh:mm:ss |
| b__ | | | Functions about probe calibration | | | | |
| | b1_ | | Probe nr. 1 | | | | |
| | | b1C | room temperature | -9.0 | 9.0 | 0.0 | K |
| | | b1A | enable probe | oFF | _on | _on | / |
| L__ | | | Functions about alarm and stand-by | | | | |
| | Lt_ | | Temperature alarm | | | | |
3 | | | LtL | low temperature alarm set point | -55.0 | 145.0 | -2.0 | °C |
4 | | | LtH | high temperature alarm set point | -55.0 | 145.0 | 14.0 | °C |
| | | Ltd | alarm delay | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | LO_ | | Door alarm | | | | |
| | | LOH | enable door alarm | oFF | _on | _on | / |
| | | LOd | door alarm delay | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | | LOt | temperature alarm minimum delay after door opening | 0 | 194 4:20:15 | 15:00 | dd hh:mm:ss |
| | Lo_ | | On / stand-by status | | | | |
| | | Loo | actual status: stand-by or on | oFF | _on | oFF | / |
| I__ | | | Functions about input-output and machine state (read only) | | | | |
| | IA_ | | Analog inputs | | | | |
| | | IA1 | room temperature | -55.0 | 145.0 | -55.0 | °C |
| | Id_ | | Digital input | | | | |
| | | Id4 | door closed | oFF | _on | oFF | / |
| | OS_ | | Machine status | | | | |
| | | OSn | fan stopped by door opening or manual control | oFF | _on | oFF | / |
| | | OSd | time left to complete the cycle | 0 | 194 4:20:15 | 0 | dd hh:mm:ss |
| | OA_ | | Analog output | | | | |
| | | LLA | actual alarm - read only (0 means no alarm) | 0 | 255 | 0 | / |
| | Od_ | | Digital output | | | | |
5 | | | Od1 | fan | oFF | _on | oFF | / |
| | | Od7 | light | oFF | _on | oFF | / |
| S__ | | | Functions about cycle | | | | |
| | St_ | | Functions about cycle temperature | | | | |
| | | _t0 | final temperature | -55.0 | 145.0 | 2.0 | °C |
1 | | | _td | final temperature differential | 0.0 | 50.0 | 0.0 | K |
| | Sd_ | | Functions about cycle timing | | | | |
2 | | | Sd0 | maximum cycle duration | 0 | 194 4:20:15 | 2:00 | dd hh:mm:ss |
| c__ | | | Functions about door and light | | | | |
| | cP_ | | Door switch and evaporator fan | | | | |
| | | cPH | stop fan when door is open | oFF | _on | _on | / |
| | | cPF | pause cycle timer when door is open | oFF | _on | _on | / |
| | | cPd | delay of fan automatic switch on | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | cI_ | | Functions about light | | | | |
| | | cIH | switch on the light when the door is open and off when closed | oFF | _on | _on | / |
| | | cIo | switch off the light automatically if it has been switched on from outside | oFF | _on | _on | / |
| | | cId | delay of light automatic switch off | 0 | 194 4:20:15 | 30 | dd hh:mm:ss |
| b__ | | | Functions about probe calibration | | | | |
| | b1_ | | Probe nr. 1 | | | | |
| | | b1C | room temperature | -9.0 | 9.0 | 0.0 | K |
| | | b1A | enable probe | oFF | _on | _on | / |
| L__ | | | Functions about alarm and stand-by | | | | |
| | Lt_ | | Temperature alarm | | | | |
3 | | | LtL | low temperature alarm set point | -55.0 | 145.0 | -2.0 | °C |
4 | | | LtH | high temperature alarm set point | -55.0 | 145.0 | 14.0 | °C |
| | | Ltd | alarm delay | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | LO_ | | Door alarm | | | | |
| | | LOH | enable door alarm | oFF | _on | _on | / |
| | | LOd | door alarm delay | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | | LOt | temperature alarm minimum delay after door opening | 0 | 194 4:20:15 | 15:00 | dd hh:mm:ss |
| | Lo_ | | On / stand-by status | | | | |
| | | Loo | actual status: stand-by or on | oFF | _on | oFF | / |
| I__ | | | Functions about input-output and machine state (read only) | | | | |
| | IA_ | | Analog inputs | | | | |
| | | IA1 | room temperature | -55.0 | 145.0 | -55.0 | °C |
| | Id_ | | Digital input | | | | |
| | | Id4 | door closed | oFF | _on | oFF | / |
| | OS_ | | Machine status | | | | |
| | | OSn | fan stopped by door opening or manual control | oFF | _on | oFF | / |
| | | OSd | time left to complete the cycle | 0 | 194 4:20:15 | 0 | dd hh:mm:ss |
| | OA_ | | Analog output | | | | |
| | | LLA | actual alarm - read only (0 means no alarm) | 0 | 255 | 0 | / |
| | Od_ | | Digital output | | | | |
5 | | | Od1 | fan | oFF | _on | oFF | / |
| | | Od7 | light | oFF | _on | oFF | / |
| S__ | | | Functions about cycle | | | | |
| | St_ | | Functions about cycle temperature | | | | |
| | | _t0 | final temperature | -55.0 | 145.0 | 2.0 | °C |
1 | | | _td | final temperature differential | 0.0 | 50.0 | 0.0 | K |
| | Sd_ | | Functions about cycle timing | | | | |
2 | | | Sd0 | maximum cycle duration | 0 | 194 4:20:15 | 2:00 | dd hh:mm:ss |
| c__ | | | Functions about door and light | | | | |
| | cP_ | | Door switch and evaporator fan | | | | |
| | | cPH | stop fan when door is open | oFF | _on | _on | / |
| | | cPF | pause cycle timer when door is open | oFF | _on | _on | / |
| | | cPd | delay of fan automatic switch on | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | cI_ | | Functions about light | | | | |
| | | cIH | switch on the light when the door is open and off when closed | oFF | _on | _on | / |
| | | cIo | switch off the light automatically if it has been switched on from outside | oFF | _on | _on | / |
| | | cId | delay of light automatic switch off | 0 | 194 4:20:15 | 30 | dd hh:mm:ss |
| b__ | | | Functions about probe calibration | | | | |
| | b1_ | | Probe nr. 1 | | | | |
| | | b1C | room temperature | -9.0 | 9.0 | 0.0 | K |
| | | b1A | enable probe | oFF | _on | _on | / |
| L__ | | | Functions about alarm and stand-by | | | | |
| | Lt_ | | Temperature alarm | | | | |
3 | | | LtL | low temperature alarm set point | -55.0 | 145.0 | -2.0 | °C |
4 | | | LtH | high temperature alarm set point | -55.0 | 145.0 | 14.0 | °C |
| | | Ltd | alarm delay | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | LO_ | | Door alarm | | | | |
| | | LOH | enable door alarm | oFF | _on | _on | / |
| | | LOd | door alarm delay | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | | LOt | temperature alarm minimum delay after door opening | 0 | 194 4:20:15 | 15:00 | dd hh:mm:ss |
| | Lo_ | | On / stand-by status | | | | |
| | | Loo | actual status: stand-by or on | oFF | _on | oFF | / |
| I__ | | | Functions about input-output and machine state (read only) | | | | |
| | IA_ | | Analog inputs | | | | |
| | | IA1 | room temperature | -55.0 | 145.0 | -55.0 | °C |
| | Id_ | | Digital input | | | | |
| | | Id4 | door closed | oFF | _on | oFF | / |
| | OS_ | | Machine status | | | | |
| | | OSn | fan stopped by door opening or manual control | oFF | _on | oFF | / |
| | | OSd | time left to complete the cycle | 0 | 194 4:20:15 | 0 | dd hh:mm:ss |
| | OA_ | | Analog output | | | | |
| | | LLA | actual alarm - read only (0 means no alarm) | 0 | 255 | 0 | / |
| | Od_ | | Digital output | | | | |
5 | | | Od1 | fan | oFF | _on | oFF | / |
| | | Od7 | light | oFF | _on | oFF | / |
| S__ | | | Functions about cycle | | | | |
| | St_ | | Functions about cycle temperature | | | | |
| | | _t0 | final temperature | -55.0 | 145.0 | 2.0 | °C |
1 | | | _td | final temperature differential | 0.0 | 50.0 | 0.0 | K |
| | Sd_ | | Functions about cycle timing | | | | |
2 | | | Sd0 | maximum cycle duration | 0 | 194 4:20:15 | 2:00 | dd hh:mm:ss |
| c__ | | | Functions about door and light | | | | |
| | cP_ | | Door switch and evaporator fan | | | | |
| | | cPH | stop fan when door is open | oFF | _on | _on | / |
| | | cPF | pause cycle timer when door is open | oFF | _on | _on | / |
| | | cPd | delay of fan automatic switch on | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | cI_ | | Functions about light | | | | |
| | | cIH | switch on the light when the door is open and off when closed | oFF | _on | _on | / |
| | | cIo | switch off the light automatically if it has been switched on from outside | oFF | _on | _on | / |
| | | cId | delay of light automatic switch off | 0 | 194 4:20:15 | 30 | dd hh:mm:ss |
| b__ | | | Functions about probe calibration | | | | |
| | b1_ | | Probe nr. 1 | | | | |
| | | b1C | room temperature | -9.0 | 9.0 | 0.0 | K |
| | | b1A | enable probe | oFF | _on | _on | / |
| L__ | | | Functions about alarm and stand-by | | | | |
| | Lt_ | | Temperature alarm | | | | |
3 | | | LtL | low temperature alarm set point | -55.0 | 145.0 | -2.0 | °C |
4 | | | LtH | high temperature alarm set point | -55.0 | 145.0 | 14.0 | °C |
| | | Ltd | alarm delay | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | LO_ | | Door alarm | | | | |
| | | LOH | enable door alarm | oFF | _on | _on | / |
| | | LOd | door alarm delay | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | | LOt | temperature alarm minimum delay after door opening | 0 | 194 4:20:15 | 15:00 | dd hh:mm:ss |
| | Lo_ | | On / stand-by status | | | | |
| | | Loo | actual status: stand-by or on | oFF | _on | oFF | / |
| I__ | | | Functions about input-output and machine state (read only) | | | | |
| | IA_ | | Analog inputs | | | | |
| | | IA1 | room temperature | -55.0 | 145.0 | -55.0 | °C |
| | Id_ | | Digital input | | | | |
| | | Id4 | door closed | oFF | _on | oFF | / |
| | OS_ | | Machine status | | | | |
| | | OSn | fan stopped by door opening or manual control | oFF | _on | oFF | / |
| | | OSd | time left to complete the cycle | 0 | 194 4:20:15 | 0 | dd hh:mm:ss |
| | OA_ | | Analog output | | | | |
| | | LLA | actual alarm - read only (0 means no alarm) | 0 | 255 | 0 | / |
| | Od_ | | Digital output | | | | |
5 | | | Od1 | fan | oFF | _on | oFF | / |
| | | Od7 | light | oFF | _on | oFF | / |
| S__ | | | Functions about cycle | | | | |
| | St_ | | Functions about cycle temperature | | | | |
| | | _t0 | final temperature | -55.0 | 145.0 | 2.0 | °C |
1 | | | _td | final temperature differential | 0.0 | 50.0 | 0.0 | K |
| | Sd_ | | Functions about cycle timing | | | | |
2 | | | Sd0 | maximum cycle duration | 0 | 194 4:20:15 | 2:00 | dd hh:mm:ss |
| c__ | | | Functions about door and light | | | | |
| | cP_ | | Door switch and evaporator fan | | | | |
| | | cPH | stop fan when door is open | oFF | _on | _on | / |
| | | cPF | pause cycle timer when door is open | oFF | _on | _on | / |
| | | cPd | delay of fan automatic switch on | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | cI_ | | Functions about light | | | | |
| | | cIH | switch on the light when the door is open and off when closed | oFF | _on | _on | / |
| | | cIo | switch off the light automatically if it has been switched on from outside | oFF | _on | _on | / |
| | | cId | delay of light automatic switch off | 0 | 194 4:20:15 | 30 | dd hh:mm:ss |
| b__ | | | Functions about probe calibration | | | | |
| | b1_ | | Probe nr. 1 | | | | |
| | | b1C | room temperature | -9.0 | 9.0 | 0.0 | K |
| | | b1A | enable probe | oFF | _on | _on | / |
| L__ | | | Functions about alarm and stand-by | | | | |
| | Lt_ | | Temperature alarm | | | | |
3 | | | LtL | low temperature alarm set point | -55.0 | 145.0 | -2.0 | °C |
4 | | | LtH | high temperature alarm set point | -55.0 | 145.0 | 14.0 | °C |
| | | Ltd | alarm delay | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | LO_ | | Door alarm | | | | |
| | | LOH | enable door alarm | oFF | _on | _on | / |
| | | LOd | door alarm delay | 0 | 194 4:20:15 | 30:00 | dd hh:mm:ss |
| | | LOt | temperature alarm minimum delay after door opening | 0 | 194 4:20:15 | 15:00 | dd hh:mm:ss |
| | Lo_ | | On / stand-by status | | | | |
| | | Loo | actual status: stand-by or on | oFF | _on | oFF | / |
| I__ | | | Functions about input-output and machine state (read only) | | | | |
| | IA_ | | Analog inputs | | | | |
| | | IA1 | room temperature | -55.0 | 145.0 | -55.0 | °C |
| | Id_ | | Digital input | | | | |
| | | Id4 | door closed | oFF | _on | oFF | / |
| | OS_ | | Machine status | | | | |
| | | OSn | fan stopped by door opening or manual control | oFF | _on | oFF | / |
| | | OSd | time left to complete the cycle | 0 | 194 4:20:15 | 0 | dd hh:mm:ss |
| | OA_ | | Analog output | | | | |
| | | LLA | actual alarm - read only (0 means no alarm) | 0 | 255 | 0 | / |
| | Od_ | | Digital output | | | | |
5 | | | Od1 | fan | oFF | _on | oFF | / |
| | | Od7 | light | oFF | _on | oFF | / |
| E__ | | | Functions about slave preferences | | | | |
| | Ed_ | | Functions about network address | | | | |
| | | EdS | slave address for local network communication | 1 | 254 | 1 | / |
| | EY_ | | Functions about display | | | | |
| | | EYY | input to show on display: 1=temperature / 2=time left in minutes | 1 | 2 | 1 | / |
| P__ | | | Functions about master preferences | | | | |
| | Pd_ | | Functions about network address | | | | |
| | | PdM | master address for global network communication | 0 | 254 | 1 | / |
| | | PdS | number of slaves connected to this master | 1 | 6 | 6 | / |